Key result
POCUS demonstrates ~41-fold higher diagnostic odds for detecting reduced LVSF across multiple assessment methods.
Why the study?
The comparative diagnostic accuracy of point-of-care ultrasound using visual qualitative assessment, E-point septal separation, and artificial intelligence-assisted assessment for detecting reduced left ventricular systolic function versus transthoracic echocardiography required evaluation.
Does point-of-care ultrasound (using VQA, EPSS, or AI-assisted assessment) accurately detect reduced left ventricular systolic function compared with transthoracic echocardiography in patients undergoing cardiac evaluation?
Meta-Analysis (n=4,647)
Does point-of-care ultrasound (using VQA, EPSS, or AI-assisted assessment) accurately detect reduced left ventricular systolic function compared with transthoracic echocardiography in patients undergoing cardiac evaluation?
Answers are grounded in this paper and cite their sources.
Visual qualitative assessment, E-point septal separation, and AI-assisted point-of-care ultrasound all provide moderate-to-high diagnostic accuracy for ruling in reduced left ventricular systolic function regardless of operator experience.
Objective To assess the diagnostic accuracy of point-of-care ultrasound (POCUS) for detecting reduced left ventricular systolic function (LVSF) using visual qualitative assessment (VQA), E-point septal separation (EPSS), and artificial intelligence (AI)-assisted assessment, compared with transthoracic echocardiography (TTE). Methods We searched PubMed, Ovid MEDLINE, Embase, and the Cochrane Library from January 1, 1985, through July 8, 2025. Eligible studies compared POCUS-assessed LVSF with left ventricular ejection fraction measured by TTE (or cardiac MRI in one study). Data were pooled using random-effects meta-analysis to determine the diagnostic odds ratio (DOR) and likelihood ratios (LR) for each of the three methods. Results Twenty-eight studies, including 4,647 participants, met inclusion criteria. Nine studies evaluating VQA yielded a DOR of 40.53 (95% CI, 15.07 to 109.00; P < 0.001) and a positive LR (LR + ) of 6.90 (95% CI, 3.75 to 12.69). Seven studies evaluating EPSS resulted in a DOR of 21.77 (95% CI, 10.25 to 46.28; P < 0.001) and a LR + of 5.70 (95% CI, 2.89 to 11.22). Five studies evaluating AI-assisted POCUS resulted in a DOR of 47.35 (95% CI, 16.87 to 132.93; P < 0.001) and a LR + of 8.37 (95% CI, 4.40 to 15.93). Meta-regression analyses revealed no statistically significant differences in diagnostic performance among the three methods. Subgroup analysis by operator experience level also showed no statistically significant difference between novice and experienced operators for each method. Conclusions VQA, EPSS, and AI-assisted POCUS all demonstrate moderate-to-high diagnostic accuracy for identifying reduced LVSF, with no significant differences among methods on meta-regression or by operator experience level. Moderately strong LR + across all three methods supports the use of POCUS for ruling in reduced LVSF. EPSS cutoffs at the lower and upper ends of the reported range showed higher sensitivity and specificity, respectively, suggesting utility as a rule-in or rule-out tool where formal echocardiography is not readily available. Registration PROSPERO CRD42024626542.
No takes yet. Share an insight, caveat, or question.
Khan et al. (2026) conducted a meta-analysis in Reduced left ventricular systolic function (LVSF) (n=4,647). Point-of-care ultrasound (POCUS) using visual qualitative assessment (VQA), E-point septal separation (EPSS), or artificial intelligence (AI)-assisted assessment vs. Transthoracic echocardiography (TTE) or cardiac MRI was evaluated on Diagnostic accuracy (diagnostic odds ratio and positive likelihood ratio) for detecting reduced left ventricular systolic function (DOR 40.53 (VQA); DOR 21.77 (EPSS); DOR 47.35 (AI-assisted), 95% CI 15.07-109.00 (VQA); 10.25-46.28 (EPSS); 16.87-132.93 (AI), p=<0.001). POCUS showed high diagnostic accuracy for reduced LVSF across VQA, EPSS, and AI-assisted methods.
Synapse has enriched 10 related papers. Consider them for comparative context: