Answers are grounded in this paper and cite their sources.
In a channel-perfused cardiac ablation phantom and conjugate heat transfer model, the larger normalized temperature reduction observed at deeper sensors is driven by a smaller baseline non-perfused temperature rise at depth and unmatched heater inputs rather than by perfusion cooling alone, with heater contact formulation representing the dominant modeling uncertainty.
No abstract yet. The journal hasn't published one we can show.
No takes yet. Share an insight, caveat, or question.
Mustafa et al. (2026) studied this question.