Key result
Low-cost ADS-VHR1 monitor correlates strongly with standard EKG for measuring heart rate variability.
Why the study?
Attention lacks objective measures in natural contexts of formal learning, prompting evaluation of whether a system recording heart rate variability from heart autonomic activity is valid and reliable.
Answers are grounded in this paper and cite their sources.
Attention is a cognitive and psychobiological variable that lacks objective measures in natural contexts of formal learning. However, it can be objectively measured from the heart autonomic activity. In the present study, the validity and reliability of a system for recording heart rate variability are demonstrated. Twelve prototypes were created and paired with standard electrocardiogram. Long segments with cardiac pulse recordings were used within ten minutes. A highly significant correlation index (p < 0.01) and a significantly high Cronbach’s alpha were obtained. In the variance analysis, all data analyzed presented a value of p < 0.05. It is concluded that there is no significant variation between paired systems; therefore, the attention detection system based on heart rate variability is valid and reliable.
No takes yet. Share an insight, caveat, or question.
Ávila et al. (2019) studied Healthy volunteers (Heart Rate Variability / Attention Detection) (n=3). Attention Detection System Based on the Variability of Heart Rate (ADS-VHR1) vs. Standard Electrocardiogram (EKG-P / LabChart) was evaluated on Correlation and agreement of RR interval measurements between ADS-VHR1 and standard EKG (R = 0.524 to 0.996; Cronbach's alpha = 0.61 to 0.99, p=<0.01). The Arduino-based ADS-VHR1 heart rate variability device correlated strongly with standard EKG across 12 prototypes.
Synapse has enriched 10 related papers. Consider them for comparative context: