Frequency-swept DFB lasers are attractive for compact wavelength-domain photonic sensing, but carrier–thermal coupling introduces sweep nonlinearity, hysteresis, and long-term wavelength drift. We demonstrate a laboratory proof-of-concept of an embedded HF-anchored swept-DFB laser correction framework for drift-resilient photonic thermometry. A compact HF gas cell provides an operational molecular wavelength anchor for sweep-to-sweep alignment and slow-drift correction, while a calibrated DFB tuning model and real-time model-inversion waveform correction are implemented on a low-power STM32 controller. Under a validated 1 Hz sawtooth sweep condition, the model-inversion correction reduces the wavelength-axis residual from approximately 142 pm to ±1.32 pm, while envelope-based repeated-sweep analysis shows reductions of 49.7% in sweep-center fluctuation and 75.7% in sweep-span fluctuation. Applied to a sapphire Fabry–Pérot temperature sensor, the corrected wavelength axis gives a conservative Allan-deviation-based temperature resolution of approximately 0.3 mK within the statistically better-supported averaging-time range. These results show that molecular-referenced embedded sweep correction can improve the stability of compact photonic thermometry systems. Further environmental, pressure, and sweep-rate validation is required before ocean deployment.