Photothermal deflection spectroscopy (PTD) measures optical absorption through the thermal response induced by a modulated pump beam. In an attenuated total reflection (ATR) configuration, the probe beam undergoes total internal reflection at the ATR–liquid interface, while the pump enters the adjacent liquid at normal incidence. Optical absorption produces refractive-index gradients and thermoelastic deformation that alter the direction of the reflected probe beam. The measured signal therefore depends on the lateral position at which the probe samples the heated region. Here we investigate this dependence using a three-dimensional finite-element model and an ATR-based PTD system. The model describes heat diffusion in the liquid and ATR crystal and calculates the resulting refractive-index gradient, surface deformation, and probe beam deflection. The calculated deflection is zero when the pump and probe beams overlap and reaches a maximum at a finite lateral offset, where the spatial gradients are largest. Measurements reproduce this behavior. For pump and probe radii of 300 μm and 150 μm, respectively, the maximum deflection occurs at an offset of approximately 280 μm. The signal also changes sign as the pump crosses the probe position. Increasing the modulation frequency reduces both the deflection amplitude and its spatial range. These results identify pump–probe offset and modulation frequency as important parameters for the alignment and interpretation of ATR-based PTD measurements.