This paper computes a detector noise budget for the instrument named in the title, from stated inputs and named formulas, and tests the result against a published value. The calculation finds that the coating temperature is 290.00 kelvin, the coating loss angle is 4.000 loss angle in units of ten to the minus fourth power, the coating thermal noise at 10.0 hertz is 32.891 zeptometres per root hertz, the coating thermal noise at 50.0 hertz is 14.709 zeptometres per root hertz. The ratio of the computed value to the published comparison value is 1.040, so the budget agrees with the published value within the stated uncertainties. The full set of computed quantities is deposited with the paper in machine-readable form, so every number in the text can be reproduced from the deposited script. Computed from stated assumptions, not measured.