Key result
Prosthetic mitral hemolysis is strongly linked to flow fragmentation, collision, or rapid jet acceleration.
Why the study?
In vitro studies showed shear stresses > 3,000 dynes/cm2 cause significant erythrocyte destruction, but whether these values occur in vivo during abnormal prosthetic mitral regurgitant flow was unknown.
Do distinct hydrodynamic patterns of regurgitant flow and elevated peak shear stress explain the occurrence of clinical hemolysis in patients with significant mitral prosthetic regurgitation?
Case-Control (n=27)
Do distinct hydrodynamic patterns of regurgitant flow and elevated peak shear stress explain the occurrence of clinical hemolysis in patients with significant mitral prosthetic regurgitation?
In mitral prosthetic regurgitation, the hydrodynamic pattern of the flow disturbance (fragmentation, collision, or rapid acceleration resulting in high shear stress) is more important than the anatomical origin of the leak (paravalvular vs central) in causing clinical hemolysis.
Answers are grounded in this paper and cite their sources.
Should not yet change prosthetic valve management; extends mechanistic insights but remains hypothesis-generating.
Objectives The aims of this study were to define the hydrodynamic mechanisms involved in the occurrence of hemolysis in prosthetic mitral valve regurgitation and to reproduce them in a numeric simulation model in order to estimate peak shear stress. Background Although in vitro studies have demonstrated that shear stresses > 3,000 dynes/cm2 are associated with significant erythrocyte destruction, it is not known whether these values can occur in vivo in conditions of abnormal prosthetic regurgitant flow. Methods We studied 27 patients undergoing reoperation for significant mitral prosthetic regurgitation, 16 with and 11 without hemolysis. We classified the origin and geometry of the regurgitant jets by using transesophageal echocardiography. By using the physical and morphologic characteristics defined, several hydrodynamic patterns were simulated numerically to determine shear rates. Results Eight (50%) of the 16 patients with hemolysis had paravalvular leaks and the other 8 had a jet with central origin, in contrast to 2 (18%) and 9 (82%), respectively, of the 11 patients without hemolysis (p = 0.12, power 0.38). Patients with hemolysis had patterns of flow fragmentation (n = 2), collision (n = 11) or rapid acceleration (n = 3), whereas those without hemolysis had either free jets (n = 7) or slow deceleration (n = 4) (p < 0.001, power 0.99). Numeric simulation demonstrated peak shear rates of 6,000, 4,500, 4,500, 925 and 950 dynes/cm2 in these five models, respectively. Conclusions The distinct patterns of regurgitant flow seen in these patients with mitral prosthetic hemolysis were associated with rapid acceleration and deceleration or high peak shear rates, or both. The nature of the flow disturbance produced by the prosthetic regurgitant lesion and the resultant increase in shear stress are more important than the site of origin of the flow disturbance in producing clinical hemolysis.
No takes yet. Share an insight, caveat, or question.
Garcia et al. (1996) conducted a case-control in Significant mitral prosthetic regurgitation (n=27). Hemolysis in prosthetic mitral valve regurgitation vs. Mitral prosthetic regurgitation without hemolysis was evaluated on Regurgitant flow patterns of fragmentation, collision, or rapid acceleration (p=< 0.001). Mitral prosthetic hemolysis was strongly associated with flow fragmentation, collision, or rapid acceleration vs free jets or slow deceleration without hemolysis.
Synapse has enriched 10 related papers. Consider them for comparative context: