for SAL. Mean external prediction-to-reference ratios were approximately 96% for CLB and 81% for SAL, distinct from controlled spike recoveries; intraday precision was 3.7% and interbatch precision was 5.8%. Against isotope-dilution LC-MS/MS, concordance correlation coefficients were 0.939 for CLB and 0.793 for SAL. A regularized multinomial logistic model achieved external balanced accuracy of 0.833, macro-F1 of 0.836 and Matthews correlation coefficient of 0.791. The reported bootstrap 95% confidence interval for balanced accuracy was 0.656-0.969, reflecting the small external set. The results support a candidate front-end triage strategy within the tested artificial matrices: differential signals identify presumptive CLB, SAL-associated high-response and mixed patterns, while any presumptive positive is referred for confirmatory LC-MS/MS rather than used for an anti-doping adjudication.