Answers are grounded in this paper and cite their sources.
Gold-nanoparticle lateral flow assays (GNP-LFAs) are widely used for point-of-care testing, but their colorimetric readout limits sensitivity. This exploratory structured review asked which signal-amplification strategies improve the limit of detection (LOD) relative to a conventional GNP-LFA, and by how much. Records came from a keyword-based, non-systematic search and were screened by one reviewer. Of 36 records assessed at full text, 29 were eligible and 26 gave a LOD ratio (51 comparator arms). Here, 19 studies (34 arms) used a gold label with a conventional comparator, and 8 (17 arms) used other labels or readouts and were summarised separately. Fold improvement ranged from no gain to about 1000-fold, and 11 of 51 arms gave less than 2-fold. Within a study, visual-limit ratios were equal to or up to about nine-fold larger than calibration-based ratios. Median gains were 162-fold for enzyme/catalytic amplification (four studies), 100-fold for physical pre-concentration (3 studies; 55-fold without the magnetic–plasmonic hybrid), 27-fold for silver/gold enhancement, and 24-fold for nanoparticle architecture. Flow control and test-zone pre-enrichment had a median of 1.4-fold (1.2-fold by calibration LOD alone; 6.3-fold by visual limit). Few studies per category, buffer-based comparisons, and non-independent arms limit these values, which do not rank mechanisms or establish clinical utility.
No takes yet. Share an insight, caveat, or question.
Mongadi et al. (2026) studied this question.