Assessing ultrasound measurements in deforming soft tissues requires material platforms that reproduce both acoustic and mechanical behavior under load. This study developed multilayer mechano‐acoustic phantoms to evaluate Dynamic Anatomical Ultrasonography (DAU) for assessing deformation of internal soft‐tissue layers, with application to the residuum of people with limb loss. Silicone‐based tissue‐mimicking materials representing skin, adipose, and muscle were acoustically characterized and assembled into phantoms with and without Nylon‐12 interlayers. The phantoms were tested under uniaxial compression with synchronized DAU, force, and video recordings, and represented by finite element (FE) models mechanically calibrated using reaction forces and external thickness changes. Minimal acoustic contrast between silicone layers limited interface detectability, whereas incorporation of interlayers improved tracking. FE models in both phantom configurations showed that surface tracking was a poor proxy for internal layer‐thickness changes. Across layers, mean absolute errors between DAU‐derived and FE‐predicted thickness reductions ranged from 0.015 to 0.17 mm in the phantom with interlayers and from 0.032 to 0.37 mm in the phantom without interlayers. These findings establish a framework for model‐referenced evaluation of DAU in multilayer soft‐tissue systems and support future in‐vitro and in ‐ vivo studies relevant to residuum assessment and prosthetic socket design.