Objective. Reliable myocardial T1 mapping is essential for quantitative cardiac tissue characterization but remains sensitive to respiratory motion, particularly during free-breathing acquisitions. This study evaluated the test–retest repeatability of MBSS-T1, a subject-specific, model-based, self-supervised deep-learning method for motion-robust myocardial T1 mapping. Approach. Seventeen healthy male volunteers aged 20–36 years underwent 3 T cardiac MRI using a 5(3)3 MOLLI acquisition under breath-hold (BH) and free-breathing (FB) conditions. Same-day repeat scans were acquired after a 20 min interval that included removal from and repositioning in the scanner. T1 maps were reconstructed using (1) MBSS-T1, (2) the contrast-agnostic deep-learning registration method SynthMorph, (3) scanner-inline motion correction, and (4) no motion correction. Repeatability was evaluated using intraclass correlation coefficients (ICC), relative Bland–Altman analysis, within-subject coefficients of variation (CVw), relative 95% repeatability coefficients (RC95), and variance-based bootstrap inference with Holm correction. Main results. All four methods included 17 complete participant-level test–retest pairs under each breathing condition. Under BH conditions, the global ICC point estimates were 0.77, 0.64, 0.65, and 0.60 for MBSS-T1, SynthMorph, scanner-inline correction, and no correction, respectively. Under FB conditions, the corresponding values were 0.41, 0.32, 0.27, and 0.16. MBSS-T1 also had the lowest descriptive CVw under both conditions. However, none of the paired comparisons of absolute test–retest differences was significant after Holm correction (all adjusted p ≥ 0.623). The smallest Holm-adjusted variance-ratio result was obtained for MBSS-T1 versus no correction under FB conditions (variance ratio 0.43, 95% CI 0.30–0.68; adjusted p = 0.058), which did not meet the prespecified significance threshold. Significance. MBSS-T1 yielded favourable descriptive repeatability estimates, particularly under FB conditions, but no pairwise comparison remained statistically significant after correction for multiple comparisons. These results characterize same-day, same-scanner measurement precision rather than absolute T1 accuracy and require confirmation in larger and more clinically diverse cohorts.
No takes yet. Share an insight, caveat, or question.
Zehavi et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: