Key result
Amplatzer Duct Occluder achieves ~93% immediate PDA closure and 100% occlusion by 24 hours.
Why the study?
Does anterograde transcatheter closure using the Amplatzer Duct Occluder achieve complete and safe closure in patients with patent ductus arteriosus?
Cohort (n=69)
Does anterograde transcatheter closure using the Amplatzer Duct Occluder achieve complete and safe closure in patients with patent ductus arteriosus?
Answers are grounded in this paper and cite their sources.
Transcatheter PDA closure with the Amplatzer Duct Occluder demonstrated 100% closure at 24 hours and intermediate-term follow-up with no complications in successfully implanted patients.
Between May 1997 and June 2000, 69 patients, ages 0.1 to 34 years, underwent attempted anterograde transcatheter closure of a patent ductus arteriosus (PDA) using the Amplatzer Duct Occluder (ADO). The ADO is a cone-shaped, self-centering, and repositionable occluder made of nitinol wire mesh. A 5Fr to 7Fr sheath was used for the delivery of the device. The mean PDA diameter (at the pulmonary end) was 4.6 +/- 1.9 mm (range 1 mm-8.5 mm). Sixty-seven of the 69 patients had successful device placement. The mean ADO smallest diameter was 6.9 +/- 1.8 mm (range 4 mm-12 mm). Complete angiographic closure occurred in 62 (92.5%) of 67 patients (95% confidence interval, 88.22%-98.77%). In five patients, there was a trivial residual shunt immediately after the procedure. At 24 hours, color Doppler flow imaging revealed complete closure in all 67 (100%) patients. The unsuccessful attempts occurred in two patients with a small, 1-mm diameter native PDA and residual PDA after surgical occlusion. Fluoroscopy time was 7.6 +/- 1.8 minutes (4 min-18 min). No complications were observed. At a median follow-up of 1.5 years (range 0.25 to 3.2 years), all patients had complete closure without complications. We conclude that transcatheter closure using the ADO is a highly effective and safe treatment for most patients with PDA.
No takes yet. Share an insight, caveat, or question.
Thanopoulos et al. (2001) conducted a cohort in Patent ductus arteriosus (PDA) (n=69). Anterograde transcatheter closure using the Amplatzer Duct Occluder (ADO) was evaluated on Complete angiographic closure immediately post-procedure (and complete closure at 24 hours and follow-up) (95% CI 88.22%-98.77%). Transcatheter PDA closure with the Amplatzer Duct Occluder achieved 92.5% immediate angiographic closure and 100% complete closure at 24 hours in 67 implanted patients.