Key result
Phase-contrast MRI matches tonometry for aortic PWV while halving inter-scan variability to ~3.4%.
Why the study?
Applanation tonometry is the clinically accepted method to quantify aortic pulse wave velocity, and a direct comparison of PWV values and reproducibility between phase contrast magnetic resonance and tonometry had not been performed.
Does phase contrast magnetic resonance combined with cross-correlation analysis improve the reproducibility of aortic pulse wave velocity measurements compared to applanation tonometry in healthy volunteers?
Observational (n=10)
Does phase contrast magnetic resonance combined with cross-correlation analysis improve the reproducibility of aortic pulse wave velocity measurements compared to applanation tonometry in healthy volunteers?
Phase contrast magnetic resonance combined with cross-correlation analysis provides aortic pulse wave velocity measurements comparable to applanation tonometry but with significantly higher inter-test reproducibility.
Answers are grounded in this paper and cite their sources.
PCMR with cross-correlation offers comparable PWV with lower variability than tonometry in volunteers; leaves open validation in patients and outcome studies.
To compare reproducibility of aortic pulse wave velocity (PWV) measurements obtained from phase contrast magnetic resonance (PCMR) and cross-correlation analysis to applanation tonometry. Increased pulse wave velocity (PWV) results from arterial stiffening and is commonly seen with ageing and in patients with atherosclerosis and/or hypertension [ 1 ]. Currently, applanation tonometry is the clinically accepted method to quantify aortic PWV and determine aortic compliance. A direct comparison of PWV values and reproducibility between PCMR and tonometry has not been done. PWV was measured in ten healthy volunteers (age: 32.9 ± 8.3 years) two times with each modality. To obtain PCMR measurements, subjects were scanned in a 1.5 T Philips Intera CV scanner (Philips Medical Systems, Best, The Netherlands). Oblique sagittal images in the plane of the aorta were acquired and velocity was encoded in the foot-head direction. After the first PCMR scan, subjects were removed from the scanner, repositioned, and the scan was repeated. After two PCMR scans, subjects remained in the supine position and underwent two measurements of the aortic (carotid-to-femoral) PWV by applanation tonometry using a Sphygmocor device (AtCor Medical, Sydney, Australia) in an adjacent room. PCMR data were analyzed by examining the velocity waveforms at multiple points along the length of the thoracic and abdominal aorta throughout the cardiac cycle. Cross correlation between flow waveforms was then used to determine the transit time between two adjacent locations along the aorta [ 2 ]. A robust bisquare-weighted linear regression algorithm was used to fit a line to the arrival time/location plot, which was then used to compute PWV. Applanation tonometry data were analyzed automatically by the Sphygmocor software using a standard transfer formula. Agreement and repeatability of PCMR and tonometry measurements were assessed by Bland-Altman statistics [ 3 ]. Inter-scan variability for each modality was evaluated with the Coefficient of Variation (CV). P-values less than 0.05 were considered statistically significant. Mean PWV values from PCMR and applanation tonometry were not statistically different (PCMR: 5.6 ± 1.1 m/s; Tonometry: 6.0 ± 1.0, P > 0.05) (Figure 1a ). The average coefficient of variation (CV) was significantly lower for PCMR compared to tonometry, 3.4% ± 2.7% versus 7.5% ± 3.2%, respectively (P < 0.05) (Figure 1b ). Bland Altman analysis showed that there was a 0.4 m/s bias with tonometry being higher and a confidence interval of ± 1.3 m/s. a) Average PWV for all subjects for both PCMR and tonometry (P > 0.05) . b) Inter-scan coefficient of variation for each modality (P < 0.05). PWV was successfully measured by PCMR combined with cross-correlation analysis and by applanation tonometry. There was no significant difference in the PWV values; however, PCMR has a statistically significant lower inter-test CV compared to tonometry. It can be concluded that PCMR and cross-correlation analysis can be combined to accurately assess PWV as an estimate of aortic compliance.
No takes yet. Share an insight, caveat, or question.
Suever et al. (2009) conducted an observational in Healthy volunteers (n=10). Phase contrast magnetic resonance (PCMR) combined with cross-correlation analysis vs. Applanation tonometry was evaluated on Inter-scan coefficient of variation (CV) of aortic pulse wave velocity (PWV) measurements (p=<0.05). Phase contrast MRI with cross-correlation yielded comparable aortic PWV to tonometry with a lower inter-scan coefficient of variation.