Answers are grounded in this paper and cite their sources.
Purpose :Automatic lesion localization can support point-of-care breast ultrasound (US), where operators often have limited training and scanners have limited compute and memory. Most segmentation models pair a pretrained encoder with a learned decoder, which adds parameters and computational cost. We propose a decoder-free soft-token segmenter that supervises each vision transformer (ViT) patch token with the fraction of lesion pixels in its patch. Methods :A layer normalization and a single linear layer map each token to one logit, and the resulting grid is upsampled bilinearly to image resolution. We evaluated three encoder sizes and two initializations on the public BUS-BRA test set, both in-domain and out-of-domain, against five fully fine-tuned models with feature pyramid network (FPN) decoders. Results :An oracle analysis showed that soft targets at patch size 8 raise the best achievable Dice to 0.970. The ViT-Tiny model, with 5.53 million parameters and a 577-parameter head, reached an in-domain Dice of 0.869 at 1,257 images/s, compared with 0.909 at 74 images/s for the most accurate FPN model DINOv2. Most of the remaining gap came from small lesions, where the model extended its mask into adjacent hypoechoic tissue, while low lesion contrast degraded all models similarly. Conclusion :These results support decoder-free soft-token segmentation as an efficient lesion localizer for resource-constrained breast US.
No takes yet. Share an insight, caveat, or question.
Raut et al. (2026) studied this question.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: