Answers are grounded in this paper and cite their sources.
This deliverable reports on calibration, characterisation, and robustness studies for a series of optical absorbance liquid filters (OALFs) with different concentration (dilution) developed in the project. It is to suggest minimal criteria in terms of physical and chemical properties of liquid absorbance filters as reference materials relevant to various types of MMDs. The project sets a target capability of characterization and calibration of optical absorbance liquid filters (OALFs) in the 0.001-3.000 absorbance interval and 220 nm-780 nm spectral interval with a standard uncertainty of 0.3 %-1.0 %. Another objective of the project has been to determine a set of minimal criteria (physical and chemical properties) for OALFs as reference materials by studies of robustness for series of absorbance liquids with different concentration (dilution). The supplied evidence shows that achieving these objectives requires efforts more than preparing nominal dilutions and various factors affect the assigned value of spectral absorbance of OALFs. Effects of measurement condition and parameters such as spectral bandwidth and wavelength scale accuracy of measurement instrument, and temperature, pH/formulation, homogeneity, container sealing, and ageing of filters have been investigated.
No takes yet. Share an insight, caveat, or question.
Manoocheri et al. (2026) studied this question.