Key result
QFR shows ~89% diagnostic accuracy and high reproducibility compared with invasive resting coronary indices.
Why the study?
Quantitative flow ratio (QFR) is an angiography-derived, adenosine- and wire-free method for functional coronary assessment, but data on its diagnostic performance and reproducibility relative to invasive indices in routine practice remained limited.
Does quantitative flow ratio provide accurate diagnostic performance and reproducibility compared to invasive physiological indices (instantaneous wave-free ratio and resting full-cycle ratio) in patients undergoing functional coronary assessment?
Cohort (n=95)
Independent observer measurement
Does quantitative flow ratio provide accurate diagnostic performance and reproducibility compared to invasive physiological indices (instantaneous wave-free ratio and resting full-cycle ratio) in patients undergoing functional coronary assessment?
Quantitative flow ratio shows good diagnostic accuracy (88.1%-89.1%) and reproducibility compared with invasive resting physiological indices, though borderline values near the 0.80 threshold may still require invasive confirmation.
Answers are grounded in this paper and cite their sources.
QFR may aid functional assessment in select vessels; leaves open diagnostic accuracy and reproducibility versus invasive indices.
Background Quantitative flow ratio is an angiography-derived, adenosine- and wire-free method for functional coronary assessment. Data on its diagnostic performance and reproducibility relative to invasive indices remain limited. Methods We prospectively enrolled 95 patients with 101 coronary vessels. Instantaneous wave-free ratio was assessed in 76 vessels and resting full-cycle ratio in 25. Haemodynamic significance was defined as an invasive index of 0.89 or less and a quantitative flow ratio of 0.80 or less. Observer 1 performed two measurements and observer 2 one independent measurement. Diagnostic performance and reproducibility were assessed using intraclass correlation coefficients, root-mean-square within-subject coefficients of variation, Bland-Altman analysis, Cohen's kappa and classification changes. Results Vessels comprised 52 left anterior descending, 19 left circumflex and 30 right coronary arteries. Correlation coefficients were 0.722, 0.649 and 0.703, respectively (all p < 0.001). Corresponding areas under the receiver-operating-characteristic curve were 0.919, 0.930 and 0.899, with diagnostic accuracies of 89.1%, 88.1% and 88.1%. Intraobserver and interobserver intraclass correlation coefficients were 0.869 and 0.855. Mean differences were 0.018 ± 0.070 and 0.016 ± 0.075, with 95% limits of agreement from −0.120 to 0.156 and from −0.131 to 0.164. Three vessels were reclassified in each comparison. Agreement did not differ significantly between vessels with invasive indices of 0.89 or less and those with values above 0.89. Conclusion Quantitative flow ratio demonstrated good diagnostic performance and reproducibility. However, individual measurement differences may be clinically relevant near the 0.80 threshold. Borderline results warrant cautious interpretation and consideration of invasive physiological confirmation when treatment decisions may be affected.
No takes yet. Share an insight, caveat, or question.
Antoniadis et al. (2026) conducted a cohort in Coronary artery disease requiring functional coronary assessment (n=95). Quantitative flow ratio (QFR) vs. Invasive physiological indices (instantaneous wave-free ratio [iFR] in 76 vessels and resting full-cycle ratio [RFR] in 25 vessels) was evaluated on Diagnostic performance (AUC and diagnostic accuracy) and reproducibility (intraclass correlation coefficients) relative to invasive physiological indices (AUC 0.919-0.930; Intra/interobserver ICC 0.869/0.855, p=<0.001). Quantitative flow ratio showed high diagnostic accuracy and reproducibility compared with invasive resting coronary indices.