Answers are grounded in this paper and cite their sources.
Before a medical vision language model is quantized or batched for deployment, someone checks that its benchmark accuracy has not changed. Equal accuracy can hide answers that did: a model that now says no to a question whose answer is yes has missed a finding, even if it gained a correct no elsewhere. We served three models, two general and one medical, under 37 stack changes that each alter one factor (weight or vision tower precision, batching and attention kernel, or the image path) on all 3,968 yes or no questions of VQA-RAD, SLAKE and PathVQA, and compared every answer with a bitwise reproducible reference. 36 of the 37 changes kept accuracy equivalent within 2 points and still flipped answers above a measured noise floor. On the medical model MedVLM-R1, 4 bit NF4 weights changed accuracy by −0.45 points [−1.39, +0.45] at 90% yet flipped 14.87% of answers [13.73, 16.00] at 95%, 511 of them from yes to no, lowering sensitivity by 10.7 points. The reference yes minus no margin predicted which answers flip (AUROC 0.887 to 0.963 across models), and abnormality questions flipped 3.12 times as often as other questions [2.17, 4.84]. Image path changes flipped no more answers than a one pixel shift of the image. A serving change should be accepted on its flip rate and sensitivity change read against a benign comparator, not on accuracy.
No takes yet. Share an insight, caveat, or question.
Sharven Rane (2026) studied this question.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: