Answers are grounded in this paper and cite their sources.
Complex micro/nano surfaces can produce locally unreliable responses and missing height data in optical measurements. This paper presents a single-scan structured-illumination dual-modal microscopy method that improves reconstruction completeness by reusing the same acquired images for two complementary reconstruction pathways. At each axial position, two phase-shifted fringe images are acquired. Two-frame differencing and one-dimensional Hilbert transform yield a computational optical-sectioning response for height reconstruction, termed CM reconstruction in this work. The first image acquired at each axial position is also used to construct an active focus variation (FV) response from local fringe carrier-band energy. No separate FV image sequence or additional spatial registration is required. The contrast-to-noise ratio of each response is used to assess reliability and guide height selection and weighting. Experiments on micro-bumps of three sizes, grinding wheel abrasive grains, and micro imprinted deep holes showed that fusion increased valid data coverage compared with either reconstruction pathway alone in all tested fields of view. Local comparisons showed that FV supplied heights where CM sidewall data were missing, while fusion retained local features from reliable CM results. Six independent measurements of a step-height standard with a nominal height of 1.0 μ m yielded a standard deviation of 0.0299 μ m in the measured step heights.
No takes yet. Share an insight, caveat, or question.
胡令 et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: