Answers are grounded in this paper and cite their sources.
Pathological chest radiographs are often assumed to be more vulnerable to channel impairments than normal ones. We simulated sending 250 NIH ChestX-ray14 radiographs from five diagnostic classes, in three payload formats sharing identical channel realizations, over a 1 km, 1550 nm free-space optical link with Gamma–Gamma fading, pointing errors and RS(255,204) coding. After an error event, diagnostic class explained 0.4–5% of the between-image variance in quality beyond grey-level statistics and JPEG size. A second, partially overlapping sample confirmed this, and a bit-level analysis explains why. For three AI readers, one uncorrectable codeword changed findings about as often as one benign image perturbation, whereas the quasi-static error events typical of millisecond fading destroyed the image and lowered the AUC by 0.14–0.46. Under retransmission, the mean delivery time of a 10 MB image rose from 40 to 43 ms with link-layer retransmission and to 233 ms with a fixed 200 ms timeout. In strong turbulence, an 8 cm aperture with a spherical-wave model lowered the SNR for 1% outage from 32.2 to 15.8 dB, relative to a plane-wave point receiver. Interleaving over a 10 MB image gained 8–22 dB at a 1 ms coherence time, but only about 2.5 dB at 10 ms. Radiograph links can be dimensioned without reference to pathology.
No takes yet. Share an insight, caveat, or question.
Zghair et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: