Key result
Transjugular Aveir leadless pacemaker implantation proves feasible with zero complications at 12 months in a toddler.
Why the study?
Pediatric pacemaker implantation carries potential complications; while epicardial systems are standard for children under 15 kg and transvenous devices carry infection and pocket erosion risks, retrievable leadless pacemakers represent a potential alternative.
Is internal jugular vein implantation of an Aveir AR2 leadless pacemaker into the right ventricle feasible and safe in a 2-year-old, 12.2-kg child with high grade AV block?
Case Report (n=1)
No
Is internal jugular vein implantation of an Aveir AR2 leadless pacemaker into the right ventricle feasible and safe in a 2-year-old, 12.2-kg child with high grade AV block?
This case report demonstrates that right internal jugular vein implantation of an Aveir AR2 leadless pacemaker into the right ventricle is feasible without complications in a 2-year-old, 12.2-kg child, maintaining stable pacing parameters at 12 months.
Answers are grounded in this paper and cite their sources.
May be feasible in select small children; leaves open need for prospective safety and efficacy data.
Pacemaker implants in pediatric patients can be filled with potential complications. Epicardial pacemaker implants are the usual mainstay for those under 15 kg and transvenous devices can have infectious and pocket erosion risk. Retrievable leadless pacemakers seem to be a potential option. Retrospective review of a 2-year-old patient with high grade AV block and overall lowering average junctional rate was performed. Internal review board approval was obtained. A 2-year-old, 12.2-kg female underwent successful right internal jugular vein implant of an Aveir AR2 into the right ventricle for backup ventricular pacing need. No complications occurred. Follow-up at 12 months demonstrated a threshold of 0.5 Volts @0.2 ms, impedance at 420 ohms and sensing at 10milliVolts with 30% pacing noted with programming at VVI 50 bpm, with predicted longevity at 20.2 years. Feasibility of internal jugular vein implant of an Aveir AR2 is possible without complication in a 2-year-old, 12.2 kg child. Further prospective studies are needed to assess true feasibility in this population.
No takes yet. Share an insight, caveat, or question.
Sharma et al. (2026) conducted a case report in High grade AV block with lowering average junctional rate (n=1). Right internal jugular vein implant of an Aveir AR2 leadless pacemaker into the right ventricle was evaluated on Procedural success, complications, and 12-month pacing parameters. Internal jugular implant of an Aveir AR2 leadless pacemaker in a 2-year-old had 0 complications and a 0.5V @0.2ms threshold at 12 months.
Synapse has enriched 10 related papers. Consider them for comparative context: