Key result
Automated BRISQUE echo quality scores correlate with patient weight and LVEF measurement bias vs CMR.
Why the study?
Standardisation of echocardiography chamber quantification faces challenges due to subjectivity during acquisition and analysis, and few studies have investigated the accuracy of echo measurements compared to a reference modality such as CMR when stratified by image quality.
Does an automated BRISQUE algorithm objectively quantify echocardiographic image quality and predict measurement bias in left ventricular functional indices compared with cardiac magnetic resonance?
Cross-Sectional (n=128)
Does an automated BRISQUE algorithm objectively quantify echocardiographic image quality and predict measurement bias in left ventricular functional indices compared with cardiac magnetic resonance?
An automated BRISQUE algorithm can objectively standardize echocardiographic image quality assessment and help predict image quality-dependent measurement biases in left ventricular functional indices like ejection fraction.
Answers are grounded in this paper and cite their sources.
May warrant caution interpreting EF in heavier patients with poorer images; leaves open prospective validation of automated quality scoring.
Background Echocardiography (echo) remains the most widely used imaging modality for the assessment, monitoring, and prognostication of the heart. Despite its prevalence, standardisation efforts for echo chamber quantification are ongoing, with challenges owing to subjectivity during acquisition and analysis. Furthermore, the confidence in derived functional indices is often dependent on the quality of the acquired images. However, few studies have investigated the accuracy of echo measurements compared to a reference modality such as cardiac magnetic resonance (CMR) imaging, when stratified by image quality. Purpose To develop an objective and automated method to quantify echo image quality, and subsequently to investigate the relationship between image quality and patient demographics, as well as the magnitude of bias in left ventricular (LV) functional indices compared with CMR. Methods Transthoracic apical 2D echo (2DE) and 3D echo (3DE) data from 128 participants (72 healthy controls and 56 patients with acquired heart disease) were used to train a BRISQUE (Blind/Referenceless Image Spatial Quality Evaluator) algorithm [1]. Briefly, feature extraction was performed by fitting pixel luminances to a generalised Gaussian distribution (GGD), followed by support vector regression to correlate features (i.e., shape, variance, and mean parameters of the GGD) to quality scores (Fig. 1). Independent BRISQUE models were trained on 580 2DE images (consisting of 2-, 3-, and 4-chamber views) and 128 targeted LV 3DE acquisitions at end-diastole, each assigned a subjective perceived quality score between 1 (poor) and 9 (excellent) by a single observer. LV indices including end-diastolic volume (EDV), end-systolic volume (ESV), ejection fraction (EF), and global longitudinal strain (GLS), were assessed according to standard guidelines. Resultant BRISQUE scores were plotted against patient demographics (age, height, weight) and the measurement bias by comparison to CMR (acquired within 1 hour of echo). Results Several linear relationships (where P-value of slope <0.05) were observed between demographics, cardiac indices, and BRISQUE scores. Increasing patient weight (and height in 3DE) were found to be associated with poorer image quality. There was no apparent relationship between image quality and age. Of interest, EF exhibited a relationship with image quality in both 2DE and 3DE (Fig. 2), whereby higher quality images tended to overestimate EF, while lower quality images underestimated EF. For 3DE, image quality dependency was also observed for ESV and GLS biases. Conclusions BRISQUE can objectively quantify image quality to produce scores which correlate to those of an expert observer, with potential utility for the standardised quantification of echo image quality. Using this method, it may be possible to predict patient characteristics which adversely impact echo quality, as well as the magnitude of measurement biases for certain functional indices. Funding Acknowledgement Type of funding sources: Public grant(s) – National budget only. Main funding source(s): Health Research Council (HRC)
No takes yet. Share an insight, caveat, or question.
Jayaneththi et al. (2022) conducted a cross-sectional in Healthy controls and patients with acquired heart disease (n=128). Automated BRISQUE (Blind/Referenceless Image Spatial Quality Evaluator) algorithm vs. Subjective perceived quality score by a single observer and CMR reference measurements was evaluated on Correlation of BRISQUE scores with expert subjective quality scores, patient demographics, and LV functional index measurement bias compared to CMR (p=<0.05). Automated BRISQUE scores correlated with patient weight and ejection fraction measurement bias in 2D and 3D echocardiography compared to CMR.