Background: Miniscrews are increasingly used in clinical practice to avoid side effects during teeth movements. The aim of the present study was to measure and compare the forces of orthogonal loads of stainless steel and titanium miniscrews with different diameters and subjected to thermal cycling. Methods: Ti-6Al-4V titanium and stainless steel orthodontic miniscrews from different manufacturers were tested. Miniscrews with diameters of 1.5 mm and 2.0 mm were evaluated. In total, 210 specimens were separated into seven groups according to manufacturer and diameter. For each group, 10 samples were tested as new, 10 specimens underwent thermocycling simulating 3 months in the oral cavity, and 10 samples underwent thermocycling simulating 6 months in the oral cavity. Specimens were positioned in an Instron Universal Testing Machine, and a shear load was applied at the neck of the miniscrew. Data were statistically analysed (p < 0.05). Results: Maximum load decreased with ageing, significantly so after six months in four titanium groups. Titanium and stainless steel miniscrews of equal diameter did not differ at any single ageing time, although titanium showed lower values when ageing times were pooled and in regression analysis (p < 0.05). Moreover, higher forces for 2.0 mm than 1.5 mm screws were necessary, both at 0.1 mm and 0.2 mm deflections and at maximum load. Conclusions: six months of thermocycling and two cycles of sterilisation seem to affect the mechanical properties of miniscrews.