Endoscopic minithoracotomy enables successful Gore septal occluder explantation with intact PFO closure and full symptom resolution.
Case Report (n=1)
No
Minimally invasive endoscopic surgical explantation of a Gore septal occluder with primary PFO closure is a feasible and effective treatment for severe device-related nickel and cobalt hypersensitivity.
Answers are grounded in this paper and cite their sources.
May offer a feasible option for device explantation in hypersensitivity; leaves open generalizability pending prospective data.
A 29-year-old female with posterior circulation transient ischaemic attacks associated with a patent foramen ovale developed persistent central and pleuritic chest pain after percutaneous closure with a 30 mm Gore septal occluder. Allergy testing identified nickel and cobalt hypersensitivity. Echocardiography showed a well-seated device without residual interatrial shunt. The patient underwent minimally invasive surgical explantation of the atrial septal occluder and primary closure of the residual patent foramen ovale. The operation was performed through a 3 cm right anterolateral minithoracotomy port with endoscopic visualization. Cardiopulmonary bypass was established using an 8 mm Dacron side graft to the right common femoral artery, percutaneous superior vena cava cannulation, and direct right femoral venous cannulation. Del Nido cardioplegia was delivered antegrade into the aortic root. A right atriotomy exposed the device. Traction on the retrieval cord (right atrial eyelet) collapsed the device into a linear configuration and allowed removal through the defect. The defect was closed primarily with running 4-0 Prolene suture. Post-bypass transoesophageal echocardiography showed an intact interatrial septum without residual shunt and preserved biventricular function. The patient was discharged on postoperative Day 5 and reported complete resolution of the presenting symptoms at follow-up.
No takes yet. Share an insight, caveat, or question.
Panchal et al. (2026) conducted a case report in Nickel and cobalt hypersensitivity with persistent chest pain after percutaneous patent foramen ovale closure with an atrial septal occluder (n=1). Minimally invasive endoscopic surgical explantation of atrial septal occluder and primary PFO closure was evaluated on Device explantation, intact interatrial septum without residual shunt, and symptom resolution. Minimally invasive endoscopic explantation of a 30 mm Gore septal occluder via a 3 cm minithoracotomy achieved intact PFO closure, discharge on Day 5, and complete symptom resolution.
Synapse has enriched 10 related papers. Consider them for comparative context: