Key result
Dual-energy RF and pulsed field VT ablation triggers VF and severe CRT-D lead dysfunction.
Case Report (n=1)
No
Dual-energy radiofrequency and pulsed field catheter ablation in a patient with a CRT-D can result in severe device dysfunction with high lead impedances requiring generator replacement, even when no intrinsic generator defect is found on post-explant analysis.
Answers are grounded in this paper and cite their sources.
Requires post-procedure device interrogation after dual-energy ablation in CRT-D patients; leaves open incidence and prevention in larger series.
No abstract yet. The journal hasn't published one we can show.
No takes yet. Share an insight, caveat, or question.
Lovrić et al. (2026) conducted a case report in Scar-related ventricular tachycardia storm with ischemic cardiomyopathy and heart failure with reduced ejection fraction (n=1). Dual-energy radiofrequency and pulsed field catheter ablation was evaluated on CRT-D device dysfunction and ventricular fibrillation induction. Dual-energy radiofrequency and pulsed field ablation for VT storm induced VF and caused high impedances on all CRT-D leads, necessitating generator replacement.