Key result
CRT-D reprogramming eliminates T-wave oversensing and phrenic stimulation while narrowing QRS by ~30 ms.
Why the study?
Does CRT-D reprogramming (adjusting ventricular sensing threshold and changing left ventricular pacing vector) resolve T-wave oversensing and symptomatic phrenic nerve stimulation in a patient with HFrEF and an implanted CRT-D?
Case Report (n=1)
No
Does CRT-D reprogramming (adjusting ventricular sensing threshold and changing left ventricular pacing vector) resolve T-wave oversensing and symptomatic phrenic nerve stimulation in a patient with HFrEF and an implanted CRT-D?
Timely recognition and non-invasive reprogramming of CRT-D parameters during routine follow-up can effectively resolve T-wave oversensing and symptomatic phrenic nerve stimulation while improving QRS duration.
Answers are grounded in this paper and cite their sources.
Reprogramming may resolve CRT-D complications in select cases; leaves open whether this extends to improved outcomes in prospective cohorts.
No abstract yet. The journal hasn't published one we can show.
No takes yet. Share an insight, caveat, or question.
Drljačić et al. (2026) conducted a case report in Heart failure with reduced ejection fraction (HFrEF, LVEF 15%) and implanted CRT-D with T-wave oversensing and phrenic nerve stimulation (n=1). CRT-D reprogramming (ventricular sensing threshold adjusted from 0.3 to 0.45 mV; LV pacing vector changed from LV1-LV2 to LV2-LV3) was evaluated on Resolution of T-wave oversensing and phrenic nerve stimulation, and change in QRS complex duration. CRT-D reprogramming eliminated T-wave oversensing and phrenic nerve stimulation while narrowing the QRS by 30 ms.
Synapse has enriched 10 related papers. Consider them for comparative context: