Answers are grounded in this paper and cite their sources.
Hip fractures caused by sideways falls impose a major health and economic burden on older adults, yet hip protectors are often rejected because they are thick, heavy, and uncomfortable. This study used finite element (FE) modeling and drop-tower testing to develop a lightweight two-zone natural rubber hip protector (TZNRHP), comprising a stiff solid-rubber inner zone over the greater trochanter and a closed-cell rubber-foam outer zone. A hyperelastic–viscoelastic FE model was validated against drop-tower tests on a surrogate hip (4.75 kg impactor, 0.52 m drop), and protectors 9–15 mm thick with six ventilation-hole patterns were evaluated (n = 5 repeated impacts; one-way ANOVA with Tukey’s test). The FE model predicted peak forces within 5.4% for all validation conditions. Thickness significantly affected peak force (p < 0.001, η2 = 0.99), and the 11 mm protector was the thinnest design significantly below the 3100 N fracture criterion, reducing the peak force from 6232 ± 160 N (unprotected) to 2785 ± 79 N (55.3% attenuation) while being 28.3% lighter than the original 15 mm design. For a female-specific threshold of 2800 N, the 12 mm design was required. Thirteen 5 mm ventilation holes (8.4% open area) raised the peak force by only 2.6% (2856 ± 73 N; not significantly different from the least-perforated pattern), and the 11 mm protector reduced the maximum femoral von Mises stress from 9.2 to 5.83 MPa. The optimized TZNRHP is a promising lightweight, ventilated hip protector for older adults.
No takes yet. Share an insight, caveat, or question.
Petdee et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: