Answers are grounded in this paper and cite their sources.
A previously reported new design of grating, adaptable for commercial soft X-ray emission spectroscopy (SXES) instrument for scanning electron microscopes (SEM), mainly for water-window energy region was fabricated and tested by installing it in an SXES spectrometer attached to SEM. The grating shows a comparable efficiency performance in the C-K and B-K emission energy region to that of the existing JS200N grating, while its efficiency performance increase with energy. The new grating shows 2.6- and 4.8-times larger efficiencies, respectively, for the first-order N-K (392 eV) and O-K (525 eV) emissions than those of existing JS200N grating, and 10-times for O-K intensity of existing JS2000 grating. The emission spectrum of a natural graphite showed not only the C-K emission spectrum but also the impurity O-K emission signal simultaneously when using this new grating. This performance will improve the applicability of SXES instrument for polymer and bio-materials, susceptible to electron beam damage.
No takes yet. Share an insight, caveat, or question.
Terauchi et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: