Key result
Pentaspline PFA achieves ~88% acute peri-mitral AFL success, but recurrent flutter remains common.
Why the study?
Despite widespread use of the pentaspline PFA catheter for atrial fibrillation, data supporting its feasibility, safety, and effectiveness in creating linear lesions to treat peri-mitral AFL were limited.
Does linear ablation using the pentaspline pulse field ablation catheter safely and effectively treat clinical peri-mitral macroreentrant atrial flutter?
Observational (n=80)
No
Does linear ablation using the pentaspline pulse field ablation catheter safely and effectively treat clinical peri-mitral macroreentrant atrial flutter?
Although the pentaspline pulse field ablation catheter achieves a high acute procedural success rate (87.5%) for peri-mitral atrial flutter, long-term efficacy is modest with a 52.5% recurrence rate due to electrical breakthrough across the mitral line.
Answers are grounded in this paper and cite their sources.
May enable PFA linear lesions for peri-mitral AFL; leaves open durability and need for prospective trials.
BACKGROUND: Despite widespread use of the pentaspline pulse field ablation (PFA) catheter to treat atrial fibrillation, there is limited data supporting its feasibility, safety, and effectiveness in creating linear lesions to treat peri-mitral atrial flutter (AFL). OBJECTIVE: To assess the safety and efficacy of using the pentaspline PFA catheter to treat peri-mitral AFL. METHODS: We performed a single-center retrospective review of consecutive patients receiving catheter ablation using the pentaspline PFA catheter to treat peri-mitral macroreentrant AFL. RESULTS: A total of 80 patients underwent catheter ablation for clinical peri-mitral AFL. Ablation was guided by electroanatomic mapping in all patients. AFL was treated with a lateral mitral line in 38 patients (47.5%), an anteroseptal mitral line in 31 patients (38.8%), a septal mitral line in 6 patients (7.5%), and an anterolateral mitral line in 5 patients (6.3%). Acute procedural success was achieved in 70 patients (87.5%) using the pentaspline catheter; the remaining 10 patients (12.5%) required additional radiofrequency ablation to achieve procedural success. Acute complications occurred in 4 (5.0%) patients: 2 (2.5%) patients sustained electrical isolation of the left atrial appendage, 1 (1.3%) had transient hematuria, and 1 (1.3%) had pericarditis. After a median follow-up of 322 days (IQR 192-432), 42 patients (52.5%) had recurrent AFL. Of these patients, 26 patients underwent a repeat ablation procedure-all demonstrated electrical breakthrough across the prior mitral line. CONCLUSIONS: While the pentaspline catheter can achieve acute procedural success when treating peri-mitral AFL with linear ablation of the mitral annulus, the long-term efficacy of this treatment strategy is modest.
No takes yet. Share an insight, caveat, or question.
Oates et al. (2026) conducted an observational in Peri-mitral macroreentrant atrial flutter (n=80). Linear ablation using pentaspline pulse field ablation (PFA) catheter was evaluated on Acute procedural success and recurrent atrial flutter at follow-up. Pentaspline PFA linear ablation for peri-mitral AFL achieved 87.5% acute success, but 52.5% of patients had recurrent AFL over a median 322-day follow-up.