Answers are grounded in this paper and cite their sources.
Background The study assessed the microbial barrier integrity of 3-way stopcocks exposed to various disinfecting agents in terms of stress cracks with potential relevance for infection prevention and control. Methods Four different 3-way stopcocks were filled with medium containing each more than 100,000 colony-forming units of Staphylococcus epidermidis, Pseudomonas aeruginosa, Escherichia coli and Candida albicans, with subsequent exposure to four different disinfecting agents for 6 and 24 h as well as for 28 days, exceeding realistic real-word exposure times. Mechanical challenge was simulated by repeated docking of syringes, combined with 24 h disinfection turns. After the disinfection terms, lacking bacterial growth from closed 3-way stopcocks as well as positive growth from the bacterial suspension inside was confirmed in brain heart infusion broth. Results Absence of bacterial growth from closed 3-way stopcocks and remaining vitality of the bacteria in the suspension within the stopcocks could be confirmed in all instances. Conclusions The 3-way stopcocks showed a robust barrier microbial integrity against long-term disinfection without and with combined mechanical challenging. The results suggest that infection prevention and control risks due to stress cracks caused by disinfection of 3-way stopcocks seems to be negligible and that their disinfection is safe to be performed.
No takes yet. Share an insight, caveat, or question.
Conrad et al. (2026) studied this question.