Why the study?
Patients with cardiac implantable electronic devices are increasingly exposed to cosmic radiation during air travel or solar radiation events, yet the effect of particle radiation on these devices had not been sufficiently quantified.
Does neutron irradiation cause cardiac implantable electronic devices (pacemakers and implantable cardioverter defibrillators) to reset to a suboptimal back-up mode?
Does neutron irradiation cause cardiac implantable electronic devices (pacemakers and implantable cardioverter defibrillators) to reset to a suboptimal back-up mode?
Answers are grounded in this paper and cite their sources.
Experimental neutron irradiation demonstrates that cosmic radiation can trigger reversible resets to back-up mode in pacemakers and ICDs, posing a low individual risk during air travel or solar events that is non-negligible at a population level.
In our aging society, the number of people relying on cardiac implantable electronic devices (CIEDs) is growing. Simultaneously, these people are increasingly exposed to cosmic radiation due to air travel or unpredictable severe solar radiation events. Single event effects (SEEs) in CIEDs have been frequently described in radiation therapy and are most likely caused by secondary neutrons. However, the effect of particle radiation on CIEDs has not been sufficiently quantified so far. Therefore, pacemakers (PMs) and implantable cardioverter defibrillators (ICDs) were irradiated with neutrons at different energies to assess their probability of resetting to a suboptimal back-up mode caused by these particles. Subsequently, the results were used to estimate the failure rate in different cosmic radiation environments, in particular on ground and at aviation altitudes. After irradiation several devices were found in back-up mode indicating the occurrence of a single event effect in the corresponding CIED. However, most of the devices could be reinitialized and reprogrammed. The rate of soft errors in CIEDs due to cosmic radiation in different scenarios in the atmospheric environment was calculated to be low for an individual patient but not negligible for the whole population.
No takes yet. Share an insight, caveat, or question.
Plettenberg et al. (2026) studied this question.