Key result
Consumer smartwatches fail to detect transient severe bradycardia or asystolic pauses within 5 minutes.
Why the study?
Smartwatches with photoplethysmography-based HR monitoring are increasingly used for arrhythmia detection, but their sensitivity for detecting transient bradyarrhythmia is poorly understood, which has implications for evaluating unexplained syncope.
Do consumer smartwatches (Apple Watch Series 10 and Samsung Galaxy Watch7) accurately detect transient bradyarrhythmia or asystolic pauses in patients undergoing electrophysiology procedures or CIED care?
Cohort (n=60)
non-randomized
Do consumer smartwatches (Apple Watch Series 10 and Samsung Galaxy Watch7) accurately detect transient bradyarrhythmia or asystolic pauses in patients undergoing electrophysiology procedures or CIED care?
Contemporary consumer smartwatches failed to detect any transient bradycardic episodes or asystolic pauses ≥4 seconds, demonstrating very limited utility in evaluating unexplained syncope.
Answers are grounded in this paper and cite their sources.
Smartwatches should not guide syncope evaluation for transient bradyarrhythmias; leaves open their diagnostic role pending larger validation studies.
Smartwatches with photoplethysmography-based heart rate (HR) monitoring are increasingly used for arrhythmia detection. However, their sensitivity for the detection of transient bradyarrhythmia is poorly understood. This may have important implications for the utility of smartwatches in the evaluation of unexplained syncope. We conducted a prospective non-randomized exploratory study of 60 patients undergoing electrophysiology procedures or cardiac implantable electronic device (CIED) care. Two commercially available smartwatches (Apple Watch Series 10 and Samsung Galaxy Watch7) were worn simultaneously by each participant. In Arm 1 (n=30), pacemaker-dependent CIED patients were temporarily reprogrammed to VVI 30 for 30 seconds to simulate transient bradycardia. In Arm 2 (n=30), patients undergoing device generator change or adenosine testing during catheter ablation wore smartwatches during clinically confirmed asystolic pauses ≥4 seconds. The primary endpoint was smartwatch detection of any HR <40 bpm within 5 minutes of the bradycardic event. The secondary endpoint was generation of an automated low HR or irregular pulse alert. In Arm 1, device reprogramming produced 30 bradycardic episodes (mean HR 31.2 ± 1.9 bpm, duration 30 seconds). In Arm 2, 42 asystolic pauses ≥4 seconds were recorded across 30 patients (mean duration 6.1 ± 1.4 seconds). Across both arms, neither smartwatch detected an HR <40 bpm within 5 minutes of any bradycardic event, and no automated low or irregular HR alerts were generated. In conclusion, contemporary smartwatches are insensitive to the detection of transient bradyarrhythmia. These findings underscore the limited utility of smartwatches in the workup of unexplained syncope.
No takes yet. Share an insight, caveat, or question.
William et al. (2026) conducted a cohort in Transient bradyarrhythmia / asystolic pauses in patients undergoing electrophysiology procedures or cardiac implantable electronic device (CIED) care (n=60). Simultaneous photoplethysmography-based heart rate monitoring using Apple Watch Series 10 and Samsung Galaxy Watch7 vs. Clinically confirmed bradycardic episodes (VVI 30 pacing for 30s or asystolic pauses ≥4s) was evaluated on Smartwatch detection of any heart rate <40 bpm within 5 minutes of the bradycardic event. Consumer smartwatches (Apple Watch Series 10 and Samsung Galaxy Watch7) detected 0% of transient bradycardic events or asystolic pauses ≥4 seconds within 5 minutes.