Why the study?
Leadless pacemakers offer a promising alternative to traditional pacemakers for complete AV block after TAVI, prompting a real-world comparison of device performance, complications, and access disparities between the two devices.
Does leadless pacemaker implantation provide similar device performance and safety compared to traditional pacemakers in patients requiring permanent pacing post-TAVI?
Does leadless pacemaker implantation provide similar device performance and safety compared to traditional pacemakers in patients requiring permanent pacing post-TAVI?
In post-TAVI patients requiring permanent pacing, leadless pacemakers demonstrated comparable adjusted device performance and a numerically lower complication rate compared to traditional pacemakers, though access disparities exist for low socioeconomic status patients.
Answers are grounded in this paper and cite their sources.
Observed unadjusted threshold difference warrants confounder adjustment before interpretation; leaves open leadless-traditional equivalence post-TAVI.
Leadless pacemakers (LPs) offer a promising alternative to traditional pacemakers (TPs) for the management of complete AV block after TAVI. This study provides a real-world comparison between TPs and LPs post-TAVI. A retrospective cohort study of 269 consecutive patients who required permanent pacemaker implantation following TAVI between 2016 and 2024 at two centers: Tulane Medical Center and Scripps Clinic were studied. Patient data was obtained from electronic medical records and included baseline demographics, TAVI procedural characteristics, zip-code based socioeconomic status, and pacemaker parameters. TP was used in 194 (72.1%) patients, and LP was used in 75 (27.9%) patients. The TP and LP cohorts had similar baseline demographics for age and sex. Before multivariable adjustment, at implantation and at 6 weeks follow-up, there was no significant difference in impedance or R wave sensing. Median RV capture threshold was statistically higher in the TP group compared to the LP group at implantation (0.50V versus 0.50V, respectively) (p=.025) and 6-week follow up (0.75V versus 0.50V, respectively) (p=.010). TP group also had a higher pacing burden (63.53% versus 31.25%, respectively) (p=.048) compared to LP which persisted at 6 weeks follow up (51.38% versus 30.74%, respectively). However, after multivariable linear regression adjusting for clinical site, comorbidities, TAVI procedural characteristics, and laboratory values, there was no statistically significant difference between the TP and LP groups in any device parameters at implantation or 6-weeks. There were 9 (3.3%) complications out of the 269 patients, all in the TP group (4.6% vs 0%) (p=.058). Lastly, low SES patients received less LP compared to high SES patients (21.1% versus 32.6%, respectively) (p=.040). LPs demonstrated similar device performance and similar safety to traditional pacemakers in post-TAVI patients.
No takes yet. Share an insight, caveat, or question.
Arshad et al. (2026) studied this question.