Answers are grounded in this paper and cite their sources.
Security measures are very important for every bio-medical device to address threats related to the patient’s safety. Due to wide-scale outsourcing, these devices are exposed to unauthorized access and can be wrongly manipulated. Wrong signal interpretation and poorly designed circuits can cause irreversible loss to the user. An amplifier is the core circuit of each bio-medical device. In this work, an amplifier is made secured against unauthorized access with help of a logic locking circuit and a secret key. This way circuit tempering and trojan insertion become more difficult, and attackers cannot analyze the real circuit behavior easily. This work presents block-level system and device-level circuits of security network and the core amplifier. The SMT algorithms to generate secret key are also presented. Correct key is required at two places in the circuit for make it accessible which brings the tampering possibility down substantially. In case of incorrect key, the amplifier performance degrades heavily and access to the bias circuit is denied. Threat model and authorization process are discussed in detail. All simulations have been carried out with Cadence Virtuoso EDA tool and 45 nm CMOS technology parameters. It is observed that gain with authorization is 28.23 dB and phase margin equals 80.13 °, whereas without authorization gain drops to 0.183 dB and phase margin becomes -12 °.
No takes yet. Share an insight, caveat, or question.
Jha et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: