In this study a comparison between direct and indirect methods of obtaining blood pressure has been made. A statistical analysis of the data supports the view that the muffling of sounds, or the fourth phase of Korotkow, bears a closer and more constant relationship to diastolic pressure measured directly. This report, therefore, recommends the acceptance of muffling of the sounds of Korotkow as the measure of diastolic pressure.