Answers are grounded in this paper and cite their sources.
Prototype padded coil improves SNR and accelerates workflows in spine imaging, indicating better patient comfort.
Motivation: Conventional posterior array (PA) coils that comprise large, low-density elements housed in plastic formers with external patient pads result in SNR degradation gaps between elements and the underlying anatomy. Goal(s): A novel, highly sensitive and conformal PA design to increase SNR and acceleration and workflow and facilitate a comfortably supported patient experience for spine and body imaging. Approach: A prototype 30-channel conformal PA was constructed using AIR™ and foam technologies to allow the body to 'comfortably sink' into the foam. Results: This novel PA demonstrated higher SNR and accommodated higher acceleration factors compared to a conventional PA. Impact: This conformal, padded posterior array design accommodates higher acceleration and SNR for in-vivo 3T spine and body imaging.
No takes yet. Share an insight, caveat, or question.
Stickle et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: