The separation and detection of six common inorganic anions (iodate (IO 3 − ), bromate (BrO 3 − ), bromide (Br − ), nitrite (NO 2 − ), nitrate (NO 3 − ), and iodide (I − )) in pure water and 35 ‰ artificial seawater were examined by ion chromatography (IC). As packing materials of separation columns, 1-aminoundecyl group chemically bonded silica (AUS) gels were prepared. Separation of the anions in pure water was achieved using separation columns (150 mm × 4.6 mm i.d.) packed with the AUS gels, 0.1 M NaCl + 5 mM phosphate buffer (pH 4.5) as eluent, and a UV detector (wavelength 225 nm). The anions in artificial seawater were separated and detected with a 300 mm-long column without interferences by matrix anions such as chloride (Cl − ) and sulfate (SO 4 2− ). The stationary phases have high-capacity anion-exchange/hydrophilic/hydrophobic interaction mixed-modes. The IC system was applied to five inorganic anions, IO 3 − , Br − , NO 2 − , NO 3 − , and I − in seawater of the Seto-Inland Sea, Japan. The detection limits (DLs, S / N = 3) were 11 µg L −1 (IO 3 − ), 93 (Br − ), 1.3 (NO 2 − ), 1.4 (NO 3 − ), and 1.1 (I − ) for a 100-µL sample injection.