Hypoxia is an important prognostic factor for head and neck cancers (HNC). Oxygen-Enhanced MRI (OE-MRI) can map hypoxia, by quantifying change in longitudinal relaxation time, T1, but the technique suffers from low signal-to-noise ratio (SNR), which reduces the sensitivity of hypoxia detection. We optimised standard T1 mapping methods on a 1.5T scanner, to select an accurate, precise, and high SNR sequence, derived in a phantom and in healthy volunteers. 3D Variable Flip Angle spoiled gradient-echo acquisition with view-sharing, flip angles 2o and 8-10o and with B1 correction applied provided suitable protocol for clinical application.
No takes yet. Share an insight, caveat, or question.
Tariq et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: