There are no randomized data to inform the extent to which transvenous cardiac leads cause tricuspid regurgitation (TR). This study sought to determine the effect of a transvenous implantable cardioverter-defibrillator (TV-ICD) on TR severity, and secondarily, on right ventricular (RV) size and function. We evaluated TR severity before and 6 months after implantable cardioverter-defibrillator insertion in a post hoc analysis of adults randomized to receive a transvenous (n = 252) or subcutaneous implantable cardioverter-defibrillator (S-ICD) (n = 251) device. TR and RV size and systolic function were assessed by echocardiographic images analyzed in a core laboratory. At baseline, at least mild TR was present in 30% of individuals. At 6 months, the proportion of participants with any TR in the TV-ICD group was 42% vs 19% in the S-ICD group (P < 0.001). The proportion with moderate or severe TR was 7% in the TV-ICD group vs 2% in the S-ICD group (P = 0.021). At 6 months, the OR of at least 1 grade worsening of TR in the TV-ICD group as compared with the S-ICD group was 7.2 (95% CI: 3.3-15.8; P < 0.001). There were no differences between groups with respect to RV size or systolic function. Six months following TV-ICD insertion, there was a 7-fold increase in the risk of at least 1 grade worsening of TR, with 7% of individuals having TR that was moderate or severe. There was no detectable difference in RV size or function; however, longer follow-up is needed.
No takes yet. Share an insight, caveat, or question.
Leong et al. (2024) studied this question.