Answers are grounded in this paper and cite their sources.
We present a validation for a novel sensor and data processing pipeline designed to measure blood glucose (BG) non-invasively using the rapid collection of a broad range of Radio Frequency (RF) waves via a decoupled antenna array. Five healthy human subjects ingested 37.5 grams of glucose solution to generate BG readings across two glycemic ranges: normoglycemic and hyperglycemic. Concurrent measurements from a Continuous Glucose Monitor (CGM) and the RF sensor were collected for comparative analysis. A Light Gradient-Boosting Machine (lightGBM) model was trained to predict BG values using 1,555 observations, where an observation is defined as data collected from 13 RF sensor sweeps paired with a single Dexcom G6® CGM value. Using this model, we predicted BG in the held-out test dataset with a Mean Absolute Relative Difference (MARD) of 12.7% in the normoglycemic range and 14.0% in the hyperglycemic range. While in early-stage validation, these results demonstrate the promise of this hardware and software technique for the non-invasive measurement of BG for practical application.
No takes yet. Share an insight, caveat, or question.
Klyve et al. (2024) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: