Answers are grounded in this paper and cite their sources.
Key Teaching Points•The automatic pacing threshold and output adjustment system is beneficial. However, it may cause life-threatening events in some patients.•Indications for automatic pacing systems should be carefully considered, especially in pacemaker-dependent patients with highly variable pacing thresholds.•The Holter electrocardiogram is essential in patients with pacemakers who are symptomatic because the pacemaker cannot detect errors on its own. •The automatic pacing threshold and output adjustment system is beneficial. However, it may cause life-threatening events in some patients.•Indications for automatic pacing systems should be carefully considered, especially in pacemaker-dependent patients with highly variable pacing thresholds.•The Holter electrocardiogram is essential in patients with pacemakers who are symptomatic because the pacemaker cannot detect errors on its own.
No takes yet. Share an insight, caveat, or question.
Nishijo et al. (2024) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: