Key result
Optimized device programming avoids S-ICD cross talk with a unipolar epicardial pacemaker by lowering pacing output.
Case Report (n=1)
Specific programming strategies, such as adjusting tachycardia detection rates and refractory periods, are crucial to prevent cross-talk and ensure appropriate arrhythmia detection in patients with both an S-ICD and a unipolar pacemaker.
Answers are grounded in this paper and cite their sources.
May warrant targeted device testing for cross-talk in combined pacemaker-ICD patients; leaves open need for standardized protocols.
Key Teaching Points•During defibrillation threshold testing in patients with a separate pacemaker system, it is important for the pacemaker to be set in VOO mode at maximum output to watch for pacer output oversensing and ventricular fibrillation undersensing by the implantable cardioverter-defibrillator (ie, test the “worst-case scenario”).•If oversensing of pacing artifact exists in the setting of complete heart block, the device cross-talk problem may be overcome if the tachycardia detection rate is set more than the double the maximum pacing rate. Lowering the pacing amplitude while maintaining an appropriate safety margin may also reduce the risk of pacer output oversensing.•A conditional zone that has a longer ventricular refractory period should be programmed in patients with paced rhythm. In addition, the programmed atrioventricular delay should be shorter than the ventricular blanking period.•During subcutaneous implantable cardioverter-defibrillator screening, both native and paced rhythms should be assessed, especially given the T-wave changes that occur during ventricular pacing as well as repolarization changes during native conduction due to T-wave memory. •During defibrillation threshold testing in patients with a separate pacemaker system, it is important for the pacemaker to be set in VOO mode at maximum output to watch for pacer output oversensing and ventricular fibrillation undersensing by the implantable cardioverter-defibrillator (ie, test the “worst-case scenario”).•If oversensing of pacing artifact exists in the setting of complete heart block, the device cross-talk problem may be overcome if the tachycardia detection rate is set more than the double the maximum pacing rate. Lowering the pacing amplitude while maintaining an appropriate safety margin may also reduce the risk of pacer output oversensing.•A conditional zone that has a longer ventricular refractory period should be programmed in patients with paced rhythm. In addition, the programmed atrioventricular delay should be shorter than the ventricular blanking period.•During subcutaneous implantable cardioverter-defibrillator screening, both native and paced rhythms should be assessed, especially given the T-wave changes that occur during ventricular pacing as well as repolarization changes during native conduction due to T-wave memory.
No takes yet. Share an insight, caveat, or question.
Kossidas et al. (2017) conducted a case report in Complete heart block and ventricular fibrillation arrests (n=1). Subcutaneous implantable cardioverter-defibrillator (S-ICD) was evaluated on Device cross talk and defibrillation efficacy. A subcutaneous implantable cardioverter-defibrillator was successfully implanted and programmed to avoid cross talk with a unipolar epicardial pacemaker by lowering pacing output.
Synapse has enriched 10 related papers. Consider them for comparative context: