Key result
Micra VR pacemaker fails suddenly with battery deformation after 4 years in a case report.
Why the study?
Leadless pacemakers rarely exhibit intrinsic device failure.
Answers are grounded in this paper and cite their sources.
Demonstrates that sudden battery failure in leadless pacemakers can occur and may be diagnosed via fluoroscopic evidence of battery deformation, emphasizing the need for ongoing monitoring.
Leadless pacemakers rarely exhibit intrinsic device failure. A Micra VR was implanted in a 68‐year‐old man with persistent atrial fibrillation and complete atrioventricular block. Four years later, despite a projected 7‐year longevity, pacing abruptly ceased and telemetry was lost. Fluoroscopy showed stable position but new battery‐pack expansion, consistent with an abrupt battery event. A Micra AV2 was implanted; high mid‐to‐lower septal thresholds improved after switching to right internal jugular access for higher septal placement. The malfunctioning Micra VR was percutaneously retrieved. This case illustrates sudden Micra VR battery failure and underscores the value of fluoroscopy, follow‐up, and remote monitoring.
No takes yet. Share an insight, caveat, or question.
Nakamura et al. (2025) studied this question. A 68-year-old man experienced sudden battery failure of the Micra VR pacemaker after 4 years, evidenced by telemetry loss and battery deformation.