Key result
Monopolar PFA transiently increases AVEIR leadless pacemaker thresholds, recovering within 24 hours.
Why the study?
Leadless atrial pacemakers are increasingly used, but their interactions with pulsed field ablation remain poorly described.
Does monopolar pulsed field ablation interfere with atrial leadless pacemakers?
Population
1 patient with an AVEIR leadless pacemaker undergoing right inferior pulmonary vein ablation
Design
Case report
Follow-up
24 h
Authors
Loading...
May warrant periprocedural monitoring of leadless pacemakers during monopolar PFA; hypothesis-generating for device interactions.
Does monopolar pulsed field ablation interfere with atrial leadless pacemakers?
Monopolar pulsed field ablation can cause transient loss of capture and pacing threshold elevation in atrial leadless pacemakers, highlighting the need for device monitoring during the procedure.
Díaz et al. (2026) studied this question. Monopolar pulsed field ablation increased the pacing threshold of an AVEIR leadless pacemaker from 0.75 V to 1.25 V, recovering within 24 hours.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: