Key result
Monopolar PFA transiently increases AVEIR leadless pacemaker thresholds, recovering within 24 hours.
Why the study?
Leadless atrial pacemakers are increasingly used, but their interactions with pulsed field ablation remain poorly described.
Does monopolar pulsed field ablation interfere with atrial leadless pacemakers?
Does monopolar pulsed field ablation interfere with atrial leadless pacemakers?
Monopolar pulsed field ablation can cause transient loss of capture and pacing threshold elevation in atrial leadless pacemakers, highlighting the need for device monitoring during the procedure.
Answers are grounded in this paper and cite their sources.
May warrant periprocedural monitoring of leadless pacemakers during monopolar PFA; hypothesis-generating for device interactions.
Introduction Leadless atrial pacemakers are increasingly used, yet interactions with pulsed field ablation (PFA) remain poorly described. Methods and Results We report the first documented case of transient loss of capture and pacing threshold elevation in an AVEIR leadless pacemaker temporally associated with monopolar PFA during right inferior pulmonary vein ablation. Threshold increased from 0.75 V @ 0.2 ms to 1.25 V @ 0.4 ms, recovering within 24 h. Conclusion Monopolar PFA can transiently interfere with atrial leadless pacing, emphasizing the need for device monitoring during ablation.
No takes yet. Share an insight, caveat, or question.
Díaz et al. (2026) studied this question. Monopolar pulsed field ablation increased the pacing threshold of an AVEIR leadless pacemaker from 0.75 V to 1.25 V, recovering within 24 hours.