Key result
Atrial leadless pacemaker linked to sub-acute pacing threshold re-elevations that normalize within 1 month.
Why the study?
Elevated acute pacing thresholds with atrial leadless pacemakers typically normalize within 24 hours, but subacute re-elevation after hospital discharge had not been well documented.
Case Report (n=1)
Sub-acute re-elevation of atrial leadless pacemaker pacing thresholds may persist for over a week but can normalize within a month, suggesting conservative management may be considered.
Answers are grounded in this paper and cite their sources.
May support expectant management of transient threshold rises; this case leaves open predictors of atrial leadless pacemaker stability.
Introduction The safety and efficacy of the atrial leadless pacemaker (A‐LP) with helix fixation have been demonstrated. Usually, elevated acute thresholds decrease gradually and normalize within 24 h. Here, we report a case of subacute re‐elevation of pacing thresholds after hospital discharge. Case Presentation An 80‐year‐old female with a history of bradycardia underwent A‐LP implantation. At the time of the procedure, the final pacing threshold (PT) was measured at 4.0 V/0.4 ms. On post‐procedural Day 1, the PT at 0.4 ms pulse width decreased to 1.0 V. On Day 4, the PT was 1.0 V/0.4 ms prior to the patient's discharge. However, on Day 5, she was readmitted to the hospital due to the loss of atrial capture at the maximum output available after an intravenous administration of 150 mg amiodarone for atrial flutter. Over 3 days of monitoring, atrial PT was higher than 6 V/1.5 ms. A traditional DDD pacemaker was implanted due to recurrent sinus asystole associated with dizziness. At the 1‐month follow‐up, the PT of atrial A‐LP was 0.5 V/0.4 ms. Two months later, an ambulatory Holter confirmed the stable and normal atrial capture of the A‐LP after lowering the base rate of the DDD pacemaker. At the 3‐month follow‐up, PT of the A‐LP remained 0.5 V/0.4 ms. Conclusion Sub‐acute re‐elevated PT of the A‐LP during the perioperative period may persist for more than 1 week and then normalize within 1 month. If PT re‐elevation is suspected to be related to peri‐electrode inflammation, a conservative approach with continued monitoring of threshold changes may be considered.
No takes yet. Share an insight, caveat, or question.
Wu et al. (2026) conducted a case report in bradycardia (n=1). Atrial leadless pacemaker (A-LP) implantation was evaluated. Atrial leadless pacemaker implantation in an 80-year-old female resulted in sub-acute re-elevated pacing thresholds that normalized to 0.5 V/0.4 ms within 1 month.
Synapse has enriched 10 related papers. Consider them for comparative context: