Key result
Combining S-ICDs and baroreflex activation therapy is feasible with appropriate testing and output optimization.
Why the study?
Baroreflex activation therapy may interfere with subcutaneous ICD sensing due to unipolar stimulation, posing a risk of arrhythmia undersensing.
Does concurrent use of baroreflex activation therapy and subcutaneous ICDs cause device interference leading to arrhythmia undersensing?
Case Report (n=1)
Does concurrent use of baroreflex activation therapy and subcutaneous ICDs cause device interference leading to arrhythmia undersensing?
Answers are grounded in this paper and cite their sources.
Concurrent use of baroreflex activation therapy and subcutaneous ICDs can cause dangerous sensing interference, requiring careful compatibility testing and output optimization.
BACKGROUND: Baroreflex activation therapy (BAT) improves symptoms and quality of life in patients with heart failure with reduced ejection fraction, while subcutaneous implantable cardioverter-defibrillators (S-ICDs) protect against sudden cardiac death without intravascular hardware. Unipolar BAT stimulation poses a theoretical risk of interference with S-ICD sensing. CASE SUMMARY: A 52-year-old man with heart failure with reduced ejection fraction, end-stage renal disease on dialysis, and an S-ICD underwent BAT implantation. During routine hemodialysis, he experienced ventricular fibrillation arrest that the S-ICD failed to detect, misclassifying the rhythm as atrial fibrillation. Device interrogation revealed high-frequency artifact corresponding to BAT output at 6 mA. BAT output was reduced to 1 mA, and S-ICD vectors were reprogrammed. Three months later, the S-ICD appropriately detected and terminated sustained monomorphic ventricular tachycardia. DISCUSSION: BAT-S-ICD interference can result in arrhythmia undersensing. With increasing adoption of both therapies, structured compatibility testing, individualized programming, and multidisciplinary follow-up are critical. TAKE-HOME MESSAGE: Concurrent S-ICD and BAT use is feasible with careful testing and output optimization.
No takes yet. Share an insight, caveat, or question.
Aigbe et al. (2026) conducted a case report in Heart failure with reduced ejection fraction (n=1). Baroreflex activation therapy (BAT) was evaluated. Concurrent use of baroreflex activation therapy and subcutaneous ICDs can cause arrhythmia undersensing due to interference, but is feasible with careful testing and output optimization.
Synapse has enriched 10 related papers. Consider them for comparative context: