Key result
The combination of a baroreflex activation therapy and a subcutaneous implantable cardioverter defibrillator was safe in a single patient, though BAT pulse width was limited by S-ICD noise detection.
Why the study?
Is the combination of a subcutaneous ICD and a baroreceptor activation device safe and feasible?
Case Report (n=1)
No
Is the combination of a subcutaneous ICD and a baroreceptor activation device safe and feasible?
This case report provides initial evidence that combining a subcutaneous ICD with a baroreflex activation therapy device is feasible and safe, provided device-device interactions are carefully managed.
Answers are grounded in this paper and cite their sources.
Requires careful programming to avoid oversensing; leaves open broader feasibility of combined BAT and S-ICD.
We present a case of a patient with a baroreflex activation therapy (BAT) receiving a subcutaneous implantable cardioverter defibrillator (S-ICD). We anticipated two possible hazardous interactions between the two devices. Stimulation by the BAT could be adjudicated as noise and result in underdetection of ventricular arrhythmias or it might be misinterpreted as ventricular arrhythmias and lead to inappropriate shocks. Postop ensing occurred, the upper limit of pulse width of the BAT was limited because of noise detection by the S-ICD, but the upper limit of amplitude was limited by patient's discomfort. In this patient, the combination of a BAT and an S-ICD was safe.
No takes yet. Share an insight, caveat, or question.
Kuffer et al. (2017) conducted a case report in Patient with a baroreflex activation therapy (BAT) receiving a subcutaneous implantable cardioverter defibrillator (S-ICD) (n=1). Combination of a baroreflex activation therapy (BAT) and a subcutaneous implantable cardioverter defibrillator (S-ICD) was evaluated on Feasibility, safety, and device interactions. The combination of a baroreflex activation therapy and a subcutaneous implantable cardioverter defibrillator was safe in a single patient, though BAT pulse width was limited by S-ICD noise detection.
Synapse has enriched 10 related papers. Consider them for comparative context: