Key result
Hydrostatic error correction raises invasive FFR from 0.78 to 0.81, improving agreement with uQFR.
Why the study?
Does correcting for hydrostatic pressure errors improve the agreement between invasive FFR and less invasive µQFR in the left anterior descending artery?
Observational (n=33)
Does correcting for hydrostatic pressure errors improve the agreement between invasive FFR and less invasive µQFR in the left anterior descending artery?
Answers are grounded in this paper and cite their sources.
Correcting for hydrostatic pressure errors in invasive FFR measurements in the LAD significantly improves agreement with angiography-derived µQFR.
Introduction/Goal Previous studies have shown a substantial agreement between the less invasive Quantitative Flow Ratio (µQFR) and the invasive Fractional Flow Reserve (FFR) measurements. However, during invasive measurements, the hydrostatic pressure error arising from the height difference between the catheter tip and the sensor could affect this concordance. Our study investigated the impact of correcting for hydrostatic pressure errors. Methods We compared 33 invasive FFR measurements performed in the left anterior descending artery (LAD) using the pressure wire with µQFR results computed from coronary angiography. Errors from hydrostatic pressure differences were corrected based on height differences measured on radiographic images (Figure 1). Results The correction of the hydrostatic error increased the FFR values from 0.78±0.11 to 0.81±0.10 (p<0.0001) resulting the same average of the µQFR values (0.81±0.10) as the corrected FFR. A strong and very similar correlation was observed between µQFR and both corrected and uncorrected invasive FFR values (r=0.954, p<0.0001 and r=0.956, p<0.0001, respectively), however the correction of the hydrostatic error has improved the agreement between the invasive FFR and the µQFR in the Blant Altmann plots (bias 0.03 and 0.00) (Figure 2). According to the ROC analysis the clinically used FFR<0.80 cut off was predicted with the <0.84 criterion of µQFR (AUC=0.976), while the hydrostatic error corrected FFR with the same cut off was associated to a <0.78 µQFR criterion (AUC=0.977). Discussion/Conclusion Our results confirm that µQFR is a reliable less invasive alternative for measuring the severity of coronary stenoses compared to invasive FFR. Hydrostatic error significantly influences the invasive FFR values in the LAD, and its correction improves the agreement between the invasive measurements and the less invasive calculations.
No takes yet. Share an insight, caveat, or question.
Piricsi et al. (2025) conducted an observational in Coronary stenosis (n=33). Hydrostatic pressure error correction for invasive FFR vs. Uncorrected invasive FFR was evaluated on Mean FFR value (p=<0.0001). Correcting for hydrostatic pressure errors increased invasive FFR values from 0.78 to 0.81, improving agreement with less invasive uQFR measurements.