February 8, 2026European Heart Journal0 citations

Posture-dependent changes in left atrial pressure: insights from the VECTOR-HF studies

View Full Paper
OKOran KorenSFS FrydmanMTM Del Trigo

Key Result

Left atrial pressure increased significantly by up to 5.61 mmHg when CHF patients changed from sitting with legs down to supine, showing posture markedly affects LAP.

Key Points

  • The analysis aimed to evaluate how left atrial pressure changes with different body postures in chronic heart failure patients.
  • Included patients from the VECTOR-HF trials.
  • Measured left atrial pressure in three postures: sitting with legs down, sitting with legs up, and supine.
  • Each posture was held for at least 3 minutes for stabilization before measurements were taken.
  • Marked variations in mean left atrial pressure associated with posture were found.
  • 69% of patients showed a clinically relevant LAP increase of more than 5 mmHg.
  • Mean LAP was lowest when sitting with legs down (12.8 mmHg) and highest in the supine position (18.4 mmHg).

Structured PICO

Does body posture significantly affect left atrial pressure measurements in chronic heart failure patients with an implanted sensor?

P
Population
13 chronic heart failure patients implanted with a left atrial pressure sensor (V-LAP) from the VECTOR-HF I, II, and IIa trials, mean age 69.1, 23% female.
I
Intervention
Left atrial pressure measurement using an implanted sensor (V-LAP) in three postures: sitting upright with legs down, sitting upright with passive leg elevation, and supine.
C
Comparator
Comparison of left atrial pressure across the three different postures.
O
Outcome
Mean left atrial pressure (LAP) value for each patient at each posture.surrogate

Body posture significantly impacts left atrial pressure measurements in heart failure patients, highlighting the need for posture-adapted targets in remote hemodynamic monitoring.

Abstract

Abstract Background Implantable hemodynamic monitoring devices were designed to collect intracardiac pressures in ambulatory heart failure (HF) patients on a day-to-day basis. Understanding posture-related hemodynamic shifts is essential for optimizing pressure-guided heart failure management. While pulmonary capillary wedge pressure, measured in the supine position, remains the standard for left-sided filling pressures, remote left atrial pressure (LAP) monitoring may provide more dynamic, posture-adapted pressure targets for ambulatory chronic heart failure (CHF) patients. The V-LAP, an interatrial device, has been proven to be safe and accurate in invasively collecting LAP in patients with HF. Patients are advised to measure LAP twice daily in a fixed posture, specifically while sitting with legs down. However, posture-related effects on LAP interpretation have not been assessed previously. Purpose This analysis aimed to evaluate left atrial pressure across various body postures using an implanted pressure sensor in CHF patients. Methods This study included patients from the VECTOR-HF I, II, and IIa trials. LAP was measured in three distinct postures: sitting upright with legs down (knees bent at a 90-degree angle), sitting upright with passive leg elevation (knees extended at a 180-degree angle, rested on a table chair), and supine (lying flat with knees extended at a 180-degree angle). Each posture was maintained for at least 3 minutes to ensure hemodynamic stabilization before obtaining multiple independent LAP measurements. The mean LAP value for each patient at each posture was then analyzed. Results A total of 13 CHF patients implanted with the LAP sensor were included (mean age 69.1 ± 10.2 years, 23% female, body mass index BMI 30.1 ± 4.2 kg/m², left ventricular ejection fraction LVEF 42.8 ± 15.3%). Variations in mean LAP were observed in all patients in relation to posture. A clinically relevant LAP increase of more than 5 mmHg was observed in 69% of patients. The mean maximum LAP change across postures was 5.8 mmHg, with the highest individual variation reaching 9.6 mmHg. Mean LAP was 12.8 ± 7.6 mmHg when sitting with legs down, 17.1 ± 9.6 mmHg when sitting with legs up, and 18.4 ± 7.2 mmHg in the supine position (Figure 1). LAP increased significantly by 5.61 mmHg (p0.001) when transitioning from sitting with legs down to supine and by 4.35 mmHg (p=0.008) from legs down to legs up. The shift from legs up to supine also showed a significant rise of 1.26 mmHg (p=0.02). Conclusions Marked variations in left atrial pressures were observed in relation to posture, particularly when transitioning from sitting with legs down to either sitting with legs up or supine. These findings highlight the significant impact of posture on LAP. Therefore, body posture should be routinely considered in HF pressure-guided management. These insights may support posture-adapted LAP target strategies for remote monitoring and optimal therapy titration.Distribution of LAP Across Postures

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Koren et al. (2025) studied this question. Left atrial pressure increased significantly by up to 5.61 mmHg when CHF patients changed from sitting with legs down to supine, showing posture markedly affects LAP.

synapsesocial.com/papers/698828210fc35cd7a884752chttps://doi.org/10.1093/eurheartj/ehaf784.1522