Key result
S-ICD shock lead looping in the device pocket causes delayed dislodgement following loop release.
Why the study?
ICD management in pediatric patients must account for small body size, growth, and disease-specific characteristics, with alternative shock lead placement strategies needed due to small vascular caliber.
Case Report (n=1)
No
Highlights a specific mechanical complication (lead dislodgement due to loop release) of subcutaneous ICD placement in pediatric patients.
Answers are grounded in this paper and cite their sources.
Calls for avoiding pocket lead loops in pediatric S-ICD; hypothesis-generating for late dislodgement prevention.
Implantable cardioverter-defibrillators (ICDs) are increasingly used in pediatric patients for both primary and secondary prevention of sudden cardiac death. In this population, ICD management requires consideration of small body size, growth-related changes, and disease-specific characteristics. In infants and young children, conventional transvenous lead placement for ICD implantation is often challenging because of the small vascular caliber, and alternative strategies—such as positioning shock leads in subcutaneous or epicardial locations—are employed on a case-by-case basis.
No takes yet. Share an insight, caveat, or question.
Kato et al. (2026) conducted a case report in Pediatric patient with congenital long QT syndrome and implanted subcutaneous ICD lead (n=1). Subcutaneous implantable cardioverter-defibrillator (ICD) shock lead replacement with loop formation in device pocket was evaluated on Lead dislodgement due to loop release in the device pocket. Subcutaneous ICD shock lead looping in the device pocket caused lead dislodgement due to loop release 50 days after implantation in a 13-year-old patient with congenital long QT syndrome.