Key result
Systematic ICD interrogation identifies lead conductor fracture behind audible alert, resolved with new lead placement.
Why the study?
ICD audible alarms cause anxiety and can stem from causes requiring urgent attention, necessitating knowledge of alert types and methodical interrogation.
Case Report (n=1)
A systematic approach to interrogating ICD audible alarms, including provocative maneuvers and impedance trends, is critical for diagnosing high voltage conductor fractures that mandate lead replacement.
Answers are grounded in this paper and cite their sources.
Systematic ICD interrogation aids lead fracture diagnosis in alarm cases; leaves open standardized protocol validation in prospective cohorts.
An audible alarm from an implantable cardioverter defibrillator (ICD) can provoke considerable anxiety in the patient. An alarm may result from several causes, some of which require urgent attention. A knowledge of the different types of audible alerts as well as methodical device interrogation is needed to unravel the cause of ICD alarms. The present case instructively illustrates the systematic approach to a patient presenting with an audible alarm from an ICD, discusses appropriate management and details of the relevant device algorithms involved.